Performance |
| Measurement ranges GMT221
GMT222 for low concentrations
Accuracy at +25 °C (+77 °F) against certified factory references (incl. repeatability and calibration uncertainty) GMT221
GMT222 Nonlinearity Sensor | 0 ... 2 % CO2 0 ... 3 % CO2 0 ... 5 % CO2 0 ... 10 % CO2 0 ... 20 % CO2
0 ... 2000 ppm 0 ... 3000 ppm 0 ... 5000 ppm 0 ... 7000 ppm 0 ... 10 000 ppm
<± [0.02% CO2 + 2 % of reading] (applies for concentration above 2% of full span) <± [20 ppm CO2 + 2 % of reading] <± 1.0% FS Vaisala CARBOCAP®
|
Operating environment |
| Temperature display Relative humidity | -20 ... +60 °C (-4 ... +140 °F) 0 ... +50 °C (+32 ... +122 °F) 0 ... 100 %RH non-condensing |
Inputs and outputs |
| Operating voltage Power consumption Analog output signals | nominal 24 VAC/VDC <4 W 0 ... 20 or 4 ... 20 mA and 0 ... 10 V |
Mechanics |
| Probe cable length Weight GMT221 GMT222 Material transmitter body probe Housing classification | 2 m and 10 m (optional)
max. 280 g max. 300 g
ABS plastic PC plastic IP65 (NEMA 4) |
Options and accessories |
| GMP221, GMP222
25245GM GM45156 25665GMSP 210848GMSP GMK220 19040GM 26150GM GM70 | spare probe (use the order form to define measurement range etc.) clips (2 pcs) for attaching the probe mounting flange for the probe 2.0 m probe cable 10.0 m probe cable calibration kit for interchangeable probes serial COM adapter for maintenance purposes chamber for pump aspirated sampling systems hand-held CO2 meter for field verification and calibration |